System Components:
The main part of the instrument;
A computer-controlled processing software;
Excitation source: quartz halogen;
Detector: PMT scans (PMT);
Heating System: Peltier thermal cycling module;
Filters: a double filter wheel.
The main parameters:
Dynamic range: 7 orders of magnitude;
Excitation range :350-750nm;
Emission range :350-700nm;
Temperature uniformity: + / -0.25 ℃;
Temperature range :29-95 ℃;
Heating and cooling rates: 2.5 ℃, seconds;
Multiple testing: simultaneous detection of up to four fluorescent dyes;
Sample volume: 96, maximum sample volume per well 200μl.
|